The zero span fill measures sweep_points RSSI samples and the following reads return them until the buffer is used up. A reader that stopped early, such as scanraw with fewer points than the sweep or an aborted sweep, left samples of its frequency and RBW behind, and the next scanraw or sweep returned them for its first points (a 29-31 MHz scanraw after a 290 point zero span scanraw showed the noise of a 600 kHz RBW at 30 MHz). The first point after a change of settings or frequency step now clears them.main