Changelog: V1.4321 ui on|off, scanraw speed-up, zero span buffer fix

main
erikkaashoek 2 weeks ago
parent 7334ad147c
commit 4823c39c06

@ -1,7 +1,9 @@
V1.4315 2026-11-12 For all tinySA Ultra models
- Console: "ui off" stops all drawing on the screen while a PC is in control, so commands no longer wait for the screen to be redrawn. The screen shows "Remote only mode. Touch screen to enable UI". A touch or a jog press gives the UI back, as do "ui on", unplugging USB, USB DISK mode and the commands that need the user (selftest, touchcal, touchtest, menu); the console then sends "ui on" before the reply to the next command. While the UI is off, messages such as errors go to the console. "ui" shows the state. It is not stored in the config or in presets and stays off when the mode changes
- STORE TINYSA NF only stores the noise figure of MEASURE TINYSA NF: after VALIDATE TINYSA NF or MEASURE AMP NF it stored a wrong value
- Leaving USB DISK mode in the signal generator shows the generator menu again, the screen stayed empty
V1.4321 2026-11-12 For all tinySA models
- Console: "ui off" stops all drawing on the screen while a PC is in control, so commands no longer wait for the screen to be redrawn (e.g. 2 ms instead of about 120 ms after a setting). The screen shows "Remote only mode. Touch screen to enable UI", "ui off "text"" adds a line under it, e.g. to tell the tracking generator from the analyzer when a PC uses two tinySA. A touch or a jog press gives the UI back, as do "ui on", unplugging USB, USB DISK mode and the commands that need the user (selftest, touchcal, touchtest, menu); the console then sends "ui on" before the next text reply. While the UI is off, messages such as errors go to the console. "ui" shows the state. It is not stored in the config or in presets and stays off when the mode changes (Ultra only)
- Console: "scanraw" no longer sets up the hardware again for every scan, only the RBW and timing of its frequency step: a one point scanraw takes about 2 ms instead of 86 ms on the tinySA basic, and the screen is no longer redrawn after each scanraw. "scanraw {f} {f} 1" (one point at one frequency) takes 2 ms instead of 93 ms on the tinySA basic and 6 ms instead of 40 ms on the Ultra
- Console: after a zero span "scanraw" with fewer points than the sweep, or an aborted zero span sweep, the next scanraw or sweep returned the remaining zero span samples for its first points, so a signal could be missing
- STORE TINYSA NF only stores the noise figure of MEASURE TINYSA NF: after VALIDATE TINYSA NF or MEASURE AMP NF it stored a wrong value (Ultra only)
- Leaving USB DISK mode in the signal generator shows the generator menu again, the screen stayed empty (Ultra only)
V1.4311 2026-11-11 For all tinySA models
- Console: "measure {mode} [value]..." starts a measurement of the MEASURE menu, the values the menu asks for are the arguments, e.g. "measure channel_power 5800M 20M". Modes: off|harmonic|oip3|phase_noise|snr|3db_width|occupied_width|am|fm|thd|channel_power|linear, on the Ultra also nf_tinysa|nf_store|nf_validate|nf_amp